Bespoke instruments are central to our research. Most of the materials we study exist only as crystals a few tens of micrometres across, or as thin films, so we design our own thermodynamic probes — for specific heat, entropy, magnetisation and length change — and run them on a dilution-refrigerator vector-magnet platform in St Andrews. This page lists the instruments developed in the group, the core platform they run on, and the techniques we use with collaborators.
Instruments developed in the group
| Technique | What it measures | Operating range and performance | Reference |
|---|---|---|---|
| Coulomb-blockade nanocalorimeter | Specific heat and magnetocaloric effect of single crystals smaller than 100 µm and of thin films | Down to 50 mK and up to 15 T; resolution of order 0.1 nJ K−1 at 500 mK on a 20 µg crystal | Cole et al., Rev. Sci. Instrum. 96, 073903 (2025) |
| A.C. elastocaloric effect under uniaxial stress (Collaboration with MPI-CPfS, Dreden / Germany) | Absolute entropy and specific heat as a function of strain, inside a piezoelectric strain cell | Dilution-refrigerator temperatures; both normal and superconducting states of Sr2RuO4 mapped | Li et al., Nature 607, 276 (2022); Hu et al., arXiv:2608.17777 (2026) |
| Piezo-driven dilatometer | Length change, thermal expansion and magnetostriction | Resolution better than 0.5 pm Hz−1/2; room temperature to a few tens of mK; fields up to 15 T | Pelly, PhD thesis, St Andrews (2026) |
| Membrane Faraday magnetometer | Magnetic moment of microcrystals and thin films, from the force on the sample in a field gradient | Membrane displacement resolved to better than 0.1 pm; sensitivity of order 10−8 emu at mK temperatures and up to 15 T | — |
| Substrate-compensated thin-film magnetometry | Paramagnetism and Curie–Weiss behaviour of thin films in a commercial SQUID magnetometer | Macor sample holder and analysis method that compensate about 99% of the substrate contribution in situ | — |
| Coulomb blockade thermometers as primary thermometers | Calibration-free, magnetic-field-independent temperature | Demonstrated from 1.5 K to 12 T in a commercial cryogen-free system | Armitage et al., Quantum Design Oxford application note (2026) |
Specific heat of ‘dust’: nanocalorimetry

Our calorimeter is built on a silicon nitride membrane and designed for microcrystals and thin films. Both the thermometer and the heater are Coulomb blockade thermometers — self-calibrating and independent of magnetic field — so the platform contains no magnetic components and can be used for both specific heat and magnetocaloric-effect measurements. Validated on a 20 µg crystal of Sr3Ru2O7, it achieves a resolution of order 0.1 nJ K−1 at 500 mK, with an absolute accuracy limited only by the sample mass. The devices are fabricated with nanofabrication partners at the Max Planck Institute for Solid State Research, Stuttgart.
Entropy under uniaxial stress
Conventional calorimetry is hard to realise inside a strain cell. We instead exploited the a.c. elastocaloric effect — the temperature oscillation of a sample under an oscillating uniaxial stress — and have developed the analysis framework that turns it into absolute entropy and specific heat across a phase diagram.
Ultrasensitive dilatometry

By combining conventional dilatometer design with the piezo-driven nanopositioning technology of scanning probe microscopy, we resolve length changes below 0.5 pm Hz−1/2 from room temperature to a few tens of millikelvin and in fields up to 15 T — enough to see the ‘breathing’ of a crystal of Sr3Ru2O7 caused by quantum oscillations.
Ultrasensitive magnetometry

Our Faraday magnetometer measures the force on a magnetic moment in a field gradient using a silicon nitride membrane whose movement is tracked to better than 0.1 pm with piezo-driven in-situ adjustment. This gives sensitivities comparable to a commercial SQUID magnetometer, around 10−8 emu, but at millikelvin temperatures and in fields up to 15 T. For thin films measured in a commercial SQUID magnetometer we developed a Macor sample holder and analysis method that compensate about 99% of the substrate contribution in situ, correcting systematic errors in quantities such as the saturation moment.
Core platform in St Andrews

The group’s main experimental platforms are a 3D vector magnet (9 T / 1 T / 1 T) that allows arbitrary magnetic-field paths in parameter space with a Kelvinox 25 dilution refrigerator, a 15T magnet with a Kelvinox 400 dilution refrigerator as well as a 14T Quantum Design PPMS and a 7T MPMS3. High-precisionmagnetoconductivity measurements are implemented in all systems.
Techniques used with collaborators
- Quantum oscillations under uniaxial stress. De Haas–van Alphen measurements under piezoelectrically applied c-axis stress on Sr2RuO4 (Jerzembeck et al., arXiv:2607.00679 (2026)).
- Cryogenic focused-ion-beam microstructuring. Transport devices with precisely defined geometry cut from selected regions of a crystal, enabling quantitative in-plane and c-axis transport on Tl2Ba2CuO6+δ (Maiti et al., arXiv:2608.02344 (2026)).
- Spectroscopic-imaging STM in vector magnetic field. A low-temperature scanning tunnelling microscope operating in a vector field, developed with the Takagi group (Zhou et al., Rev. Sci. Instrum. 94, 033704 (2023)).
Shared facilities
St Andrews has a strong condensed-matter community whose facilities we use: the thin-film growth capability of the Centre for Designer Quantum Materials, clean-room and sample-characterisation facilities within the School of Physics and Astronomy, and the School’s shared materials-measurement suite. Through the Scottish Universities Physics Alliance (SUPA) and PHYESTA we also have access to the Quantum Design PPMS at the Centre for Science at Extreme Conditions (CSEC), University of Edinburgh.